产品介绍
特性
- Accuracy: ± 1% of reading (when the test current is > 500μ / impedance in Ohms and > 30 mA)
- Range: 0 to 100 Ohms
- Measurement Technique: Successive approximation analog to digital conversion of sensed voltage and current
- Measurement window: Either 2 or 3 full sine waves depending upon line frequency (e.g. 3 at 60 Hz, 2 at 50 Hz)
- Sample rate: ~ 53 KHz, yielding ~ 2.6K voltage and 2.6K current samples per measurement window at 60 Hz, and ~ 2.1K voltage and 2.1K current samples/window at 50 Hz
- Mathematical conversion to true rms voltage and current values, division independent of phase to yield impedance
- Display derived from four layer stack averaging
- Voltage is sensed at the probe heads by internal probe sense lines or, for Model 109 probes, by precision Kelvin sensing
- Accuracy: ± 1% of reading ± 4 mA
- Range: 0 to 144 Amps rms
- Accuracy: ± 1% of reading ± 4 mA at transition points (every 2.5 Arms)
- Range: 2.5 to 77.5 Amps rms (0 to 2.5 Arms implied) (Note 3, Graph 2)
- Measurement Technique: Logical translation of digital current measurement
- Accuracy: ± .1% (Digital Timer and Audio / Visual Alerts applicable to Model 309 or, if server driven, Model 308)
- Range: 1 second to 59 minutes, 59 seconds (extended times possible under server software)
- Internal Reference: 16.777 MHz (2e24) crystal controlled system clock
- When enabled, sounds and flashes when any user specified test current or impedance limit is violated
- Accuracy: Logically detected. The accuracy therefore exactly matches the accuracy of associated measurements
- Continuous to test currents of 35 Amps
- Derate linearly to 50% duty cycle at 60 Amps
- Maximum Cycle Period: 35 to 60 Amps: 4 Minutes
- Maximum allowed common mode voltage to ground: 50 Vrms, 71 Vpeak or dc
- Maximum normal mode open circuit probe source voltage (at 132 Vrms line voltage): 8.6 Vrms, 12.1 Vpeak
- 90 to 132 Volts rms
- 48 to 66 Hertz sine wave
- Maximum Power: 396 Watts (3.0 Arms at 132 Vrms)
- Resistance: 100 milliohms ± 1%
- Maximum continuous current: 17.5 Amps
- There are no hardware calibration adjustments, and thus no hardware calibration is required or possible
- Software calibration data is available, but it's normally unnecessary to modify this data
- Complete the formal performance verification procedure annually or following any repair service
- Designed and manufactured to meet or exceed UL 3111
- Instruments manufactured before January 2008 were so listed by ETL
- Operating Temperature Range: 5 to 40°C
- Relative Humidity Range: 0 to 90% non-condensing
- Height: 10.92 centimeters (4.30 inches)
- Width: 21.15 centimeters (8.33 inches)
- Depth: 26.0centimeters (10.24 inches)
"The 309 is a Milliohm Meter from Hypatia. When engineers need to collect information regarding electrical connections such as wires, conduits, or cables, they use a milliohm meter. This piece of electronic testing equipment measures the smallest resistance values. An engineer would use a milliohm meter instead of a traditional ohmmeter to document and log resistance amounts at the lower end of the scale (small values of ohms).","Additional Features:","Electrical","Impedance Measurement","Digital Current Measurement","Bar Graph Current Measurement","Digital Timer","Audio\/Visual Alerts","Operating Duty Cycle","Input Power Requirements","Check Resistor","Verification Schedule","Safety Certification","Environmental\\","Physical","Overall Dimensions","The Hypatia 309 is a High Current Sourcing Milliohm Meter."
产品替代
资料下载
文件名称 | 大小 | 操作 |
---|---|---|
Hypatia 308, 309 | 39.57 KB | 下载 |
声明:本产品内容及配图源自互联网收集或平台用户自行上传,目的在于传递更多信息,并不代表本网赞同其观点或证实其内容真实性,不承担此类作品侵权行为的直接责任及连带责任。如涉及作品内容、版权等问题,请联系本网处理,侵权内容将在一周内下架整改。
毫欧表相关品牌
毫欧表同类产品
加载中···